EasyManua.ls Logo

RADIOMETER ABL800 FLEX - Page 87

RADIOMETER ABL800 FLEX
438 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ABL800 FLEX Operator's Manual 3. Installation and setup
Analysis setup, Continued
Setting
reference and
critical limits
(continued)
Step Action
3.
To remove all current ranges, press the Clear limits button.
4.
To change the value in a range limit, touch and highlight the range.
Then enter the limit, using the screen keyboard and confirm with
Enter.
5.
Repeat for other limits in a similar way.
Press Back when completed.
6.
Reportable
ranges
To change or enter a reportable range (must be selected smaller than or equal to the
measuring range), do the following:
Step Action
1.
Scroll to the desired parameter, using the up/down arrows or scroll
bar.
2.
Key in the desired lower limit and confirm with Enter on the keypad.
3.
Key in the upper limit and confirm with Enter on the keypad.
Measured
parameters show
their measuring
ranges. Derived
parameters show
.....
4.
To change the reportable range to the default (primary) setting,
highlight the desired parameter and press Set Default.
Continued on next page
3-21

Table of Contents

Other manuals for RADIOMETER ABL800 FLEX

Related product manuals