3. Installation and setup ABL800 FLEX Operator's Manual
Analysis setup, Continued
Reportable
ranges
(continued)
Step Action
5.
To change all parameters to the default values, press the 1-n Set All
Default button.
Press Continue to change all parameters’ reportable ranges to the
default ones.
Press Cancel to keep the user-defined reportable ranges and return to
the previous screen.
Press Close to exit the program and confirm the selected settings.
6.
See also Calibration verification in chapter 15 in this manual.
Sample Pre-
registration
setup
This program allows you to select interpretation of the barcode and the patient data
that can be confirmed before a sample measurement.
To select the settings, do the following:
Step Action
1.
Use the up/down arrow buttons to select the interpretation of the
barcode setting in the "Interpret barcode input as" box.
Choose one of the following:
Patient ID
Accession Number
Sampler ID.
Continued on next page
3-22