Common Measurement Settings
R&S
®
ESW
159User Manual 1177.6298.02 ─ 08
Overloading
At a reference level of -10 dBm and at an external generator output level of the same
value, the R&S ESW operates without overrange reserve. That means the R&S ESW
is in danger of being overloaded if a signal is applied whose amplitude is higher than
the reference line. In this case, either the message "RF OVLD" for overload or "IF
OVLD" for exceeded display range (clipping of the trace at the upper diagram border =
overrange) is displayed in the status line.
Overloading can be avoided as follows:
Reducing the output level of the external generator ("Source Power" on page 161
in "External Generator > Measurement Configuration")
Increasing attenuation in the "Amplitude" menu ("Attenuation" on page 121).
10.6.3.2 External Generator Control Settings
Access: [INPUT/OUPUT] > "External Generator Config"
The "External Generator" settings are available if the R&S ESW External Generator
Control option is installed. For each measurement channel, you can configure one
external generator. To switch between different configurations, define multiple mea-
surement channels.
For more information on external generator control, see Chapter 10.6.3.1, "Basics on
External Generator Control", on page 151.
Interface Configuration Settings............................................................................159
Measurement Settings.......................................................................................... 161
Interface Configuration Settings
Access: "Overview" > "Input" > "External Generator" > "Interface Configuration"
Generator Type........................................................................................................... 160
Interface...................................................................................................................... 160
TTL Handshake...........................................................................................................160
GPIB Address/TCPIP Address / Computer Name...................................................... 160
Reference....................................................................................................................160
Edit Generator Setup File............................................................................................160
Frequency Min/ Frequency Max..................................................................................161
Level Min/ Level Max.................................................................................................. 161