Common Measurement Settings
R&S
®
ESW
164User Manual 1177.6298.02 ─ 08
For more information about the digital baseband output refer to the documentation of
the I/Q Analyzer.
Output Coupling.......................................................................................................... 164
Selecting the output type.............................................................................................164
Configuring the output of the IF signal..........................................................164
Configuring the output of the Video signal....................................................165
Configuring the output of an AM signal.........................................................165
Configuring the output of an FM signal.........................................................165
Controlling and configuring the output........................................................................ 165
IF Output Frequency.....................................................................................165
Coupling........................................................................................................165
Reference Level for Output...........................................................................165
Low Pass...................................................................................................... 165
Phones..........................................................................................................166
Squelch.........................................................................................................166
Scale.............................................................................................................166
Controlling the volume................................................................................................ 166
Output Coupling
Selects the scope of the output settings.
"Global"
The output settings apply to all measurement channels / applications.
"Channel Spe-
cific"
The output settings apply to the current measurement channel / appli-
cation only. You can configure each channel separately.
Remote command:
OUTPut<ou>:LINK on page 499
Selecting the output type
Selects the type of analog signal you want to output.
"IF Output"
Outputs the IF signal (see "Configuring the output of the IF signal"
on page 164 for available settings).
(Unavailable for audio output.)
"Video"
Outputs the video signal (see "Configuring the output of the Video
signal" on page 165 for available settings).
"AM"
Outputs the AM demodulated signal (see "Configuring the output of
an AM signal" on page 165 for available settings).
"FM"
Outputs the FM demodulated signal (see "Configuring the output of
an FM signal" on page 165 for available settings).
"Current
Focus"
Outputs the data of the currently selected measurement window
(highlighted with a blue frame).
Available in the analog demodulation application.
"Off"
Turns off the output.
Remote command:
OUTPut<ou>:IF[:SOURce] on page 498
Configuring the output of the IF signal ← Selecting the output type