Common Analysis and Display Functions
R&S
®
ESW
245User Manual 1177.6298.02 ─ 08
Storing and Recalling Limit Lines
Limit lines can be stored with the configuration settings so they can be recalled for
other measurements at a later time (see Chapter 12.3, "Storing and Recalling Instru-
ment Settings and Measurement Data", on page 260). Note, however, that any
changes made to the limit lines after storing the configuration file cannot be restored
and will be overwritten by the stored values when the configuration file is recalled.
Always remember to store the settings again after changing the limit line values.
After recalling measurement settings, the limit line values applied to the measurement
may be different to those displayed in the "Limit Lines" dialog box; see "Saving and
recalling transducer and limit line settings" on page 261.
11.5.2.2 Limit Line Settings and Functions
Access: "Overview" > "Analysis" > "Lines"
Up to 8 limit lines can be displayed simultaneously in the R&S ESW. Many more can
be stored on the instrument.
Stored limit line settings
When storing and recalling limit line settings, consider the information provided in "Sav-
ing and recalling transducer and limit line settings" on page 261.
Limit Line Management.........................................................................................245
Limit Line Details...................................................................................................248
Limit Line Management
Access: "Overview" > "Analysis" > "Lines" > "Limit Lines"
Display and Limit Lines

Table of Contents