EasyManua.ls Logo

Rohde & Schwarz ESW Series - Page 440

Rohde & Schwarz ESW Series
759 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Commands in the Receiver Application
R&S
®
ESW
440User Manual 1177.6298.02 ─ 08
To prevent overlapping execution of the subsequent command before the measure-
ment has been aborted successfully, use the *OPC or *WAI command after ABORt and
before the next command.
To abort a sequence of measurements by the Sequencer, use the INITiate:
SEQuencer:ABORt command.
Note on blocked remote control programs:
If a sequential command cannot be completed, for example because a triggered sweep
never receives a trigger, the remote control program will never finish. The remote chan-
nel to the R&S ESW is blocked for further commands. In this case, you must interrupt
processing on the remote channel first to abort the measurement.
To do so, send a "Device Clear" command from the control instrument to the
R&S ESW on a parallel channel to clear all currently active remote channels. Depend-
ing on the used interface and protocol, send the following commands:
Visa: viClear()
GPIB: ibclr()
RSIB: RSDLLibclr()
Now you can send the ABORt command on the remote channel that runs the measure-
ment.
Example:
ABOR;INIT:IMM
Aborts the measurement and restarts it.
Usage: Event
HOLD
This command interrupts (holds) a scan.
To resume the scan, use INITiate<mt>[:IMMediate] on page 442.
Example:
HOLD
Interrupts the scan.
Usage: Event
INITiate<mt>:CONMeas
This command resumes a scan that was interrupted by a transducer stop at the current
receiver frequency.
If the scan was interrupted by the HOLD command, you have to resume it with
INITiate<mt>[:IMMediate].
Suffix:
<mt>
.
INITiate1 is irrelevant.
INITiate2 resumes the scan.
Measurements and Result Displays