Remote Commands in the Receiver Application
R&S
®
ESW
466User Manual 1177.6298.02 ─ 08
Parameters:
<Peaks> Number of peaks in one subrange.
Note that the maximum number of peaks is 500. Thus, the maxi-
mum number of peaks per subrange depends on the number of
subranges you have defined.
*RST: 1
Example:
CALC:PEAK:METH SUBR
CALC:PEAK:SUBR 20
CALC:PEAK:SUBR:PCO 5
Looks for 5 peaks in each of 20 subranges.
Manual operation: See "Controlling the size of the peak list" on page 125
CALCulate<n>:PSEarch:SUBRanges <Peaks> | <Subranges>
CALCulate<n>:PEAKsearch:SUBRanges[:VALue] <Peaks | Subranges>
The effects of this command depend on the peak search mode that you have selected.
Suffix:
<n>
.
irrelevant
Parameters:
<Peaks | Subranges> If you have selected the "Peaks" search mode, the command
defines the number of peaks to look for during the peak search.
Range: 1 to 500
*RST: 50
Example: //Select subrange search and define 25 subranges
CALC:PEAK:METH SUBR
CALC:PEAK:SUBR 25
Manual operation: See "Controlling the size of the peak list" on page 125
DISPlay[:WINDow<n>]:TRACe<t>:SYMBol <Symbol>
This command turns the peak labels in the diagram on and off.
Suffix:
<n>
.
irrelevant
<t> irrelevant
Parameters:
<Symbol> CROSs
Each peak is labeled by a symbol. The symbol and its color
depend on the trace the peak is on.
OFF
Peak labels are off.
*RST: CROSs
Example: //Turn on peak labels
DISP:TRAC:SYMB CROS