EasyManua.ls Logo

Rohde & Schwarz ESW Series - Page 465

Rohde & Schwarz ESW Series
759 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Commands in the Receiver Application
R&S
®
ESW
465User Manual 1177.6298.02 ─ 08
CALCulate<n>:PSEarch:MARGin <Margin>
CALCulate<n>:PEAKsearch:MARGin <Margin>
This command defines the limit line margin that is considered during a peak search.
Suffix:
<n>
.
irrelevant
Parameters:
<Margin> Default unit: dB
Example: //Define a limit line margin
CALC:PEAK:MARG 2
Manual operation: See "Defining peak characteristics" on page 126
CALCulate<n>:PSEarch:METHod <Method>
CALCulate<n>:PEAKsearch:METHod <Method>
This command selects the way the R&S ESW creates a peak list.
Suffix:
<n>
.
irrelevant
Parameters:
<Method> SUBRange
Divides the scan range into smaller subranges and looks for a
particular number of peaks in each subrange.
PEAK
Looks for a particular number of peaks over the complete scan
range.
*RST: PEAK
Example: //Divide the scan range into smaller subranges for the peak
search
CALC:PEAK:METH SUBR
Manual operation: See "Selecting the peak search method" on page 125
CALCulate<n>:PSEarch:SUBRanges:PCOunt <Peaks>
CALCulate<n>:PEAKsearch:SUBRanges:PCOunt <Peaks>
This command defines the number of peaks to be found in each subrange.
Prerequisites for this command
Select subrange peak search mode (CALCulate<n>:PEAKsearch:METHod).
Define number of subranges (CALCulate<n>:PEAKsearch:SUBRanges[:
VALue]).
Suffix:
<n>
.
irrelevant
Measurements and Result Displays

Table of Contents

Related product manuals