Remote Commands in the Receiver Application
R&S
®
ESW
544User Manual 1177.6298.02 ─ 08
CALCulate<n>:MARKer<m>:X:SLIMits:LEFT <SearchLimit>
This command defines the left limit of the marker search range for all markers in all
windows.
Suffix:
<n>
.
irrelevant
<m> irrelevant
Parameters:
<SearchLimit> The value range depends on the frequency range or measure-
ment time.
The unit is Hz for frequency domain measurements and s for
time domain measurements.
*RST: left diagram border
Default unit: HZ
Example:
CALC:MARK:X:SLIM ON
Switches the search limit function on.
CALC:MARK:X:SLIM:LEFT 10MHz
Sets the left limit of the search range to 10 MHz.
Manual operation: See "Search Limits (Left / Right)" on page 235
CALCulate<n>:MARKer<m>:X:SLIMits:RIGHt <SearchLimit>
This command defines the right limit of the marker search range for all markers in all
windows.
Suffix:
<n>
.
irrelevant
<m> irrelevant
Example:
CALC:MARK:X:SLIM ON
Switches the search limit function on.
CALC:MARK:X:SLIM:RIGH 20MHz
Sets the right limit of the search range to 20 MHz.
Manual operation: See "Search Limits (Left / Right)" on page 235
CALCulate<n>:THReshold <Level>
This command defines a threshold level for the marker peak search (for all markers in
all windows).
Suffix:
<n>
.
irrelevant
Parameters:
<Level> Numeric value. The value range and unit are variable.
*RST: -120 dBm
Default unit: DBM