Remote Commands in the Receiver Application
R&S
®
ESW
562User Manual 1177.6298.02 ─ 08
ON | 1
Switches the function on
Example:
CALC:FLIN2:STAT ON
Turns frequency line 2 on.
CALCulate<n>:TFLine:STATe <State>
This command turns the frequency line representing the current receiver frequency on
and off.
Suffix:
<n>
.
irrelevant
Parameters:
<State> ON | OFF | 1 | 0
*RST: ON
Example: //Turn on the frequency line
CALC:TFL:STAT ON
Manual operation: See "Tuned Frequency" on page 240
CALCulate<n>:TLINe<dl> <Time>
This command defines the position of a time line.
Suffix:
<n>
.
Window
<dl> 1 | 2
time line
Parameters:
<Time> Note that you can not set a time line to a position that is higher
than the current sweep time.
Range: 0 s to 1600 s
*RST: (STATe to OFF)
Default unit: S
Example:
CALC:TLIN 10ms
Sets the first time line to 10 ms.
Manual operation: See "Vertical Line <x>" on page 240
CALCulate<n>:TLINe<dl>:STATe <State>
This command turns a time line on and off
Suffix:
<n>
.
Window
Analysis