Remote Commands in the Receiver Application
R&S
®
ESW
566User Manual 1177.6298.02 ─ 08
<li> Limit line
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:LIM:STAT ON
Switches on the limit check for limit line 1.
Manual operation: See "Disable All Lines" on page 248
CALCulate<n>:LIMit<li>:TRACe<t> <TraceNumber>
This command links a limit line to one or more traces.
Note that this command is maintained for compatibility reasons only. Limit lines no lon-
ger need to be assigned to a trace explicitly. The trace to be checked can be defined
directly (as a suffix) in the new command to activate the limit check (see
CALCulate<n>:LIMit<li>:TRACe<t>:CHECk on page 566).
Suffix:
<n>
.
Window
<li> Limit line
<t> irrelevant
Example:
CALC:LIM2:TRAC 3
Assigns limit line 2 to trace 3.
CALCulate<n>:LIMit<li>:TRACe<t>:CHECk <State>
This command turns the limit check for a specific trace on and off.
To query the limit check result, use CALCulate<n>:LIMit<li>:FAIL?.
Note that this command replaces the two commands from previous signal and spec-
trum analyzers (which are still supported, however):
CALCulate<n>:LIMit<li>:TRACe<t> on page 566
CALCulate<n>:LIMit<li>:STATe on page 565
Suffix:
<n>
.
Window
<li> Limit line
<t> Trace
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
Analysis