Remote Commands in the Receiver Application
R&S
®
ESW
569User Manual 1177.6298.02 ─ 08
<li> Limit line
Parameters:
<Comment> String containing the description of the limit line.
Manual operation: See "Comment" on page 249
CALCulate<n>:LIMit<li>:CONTrol[:DATA] <LimitLinePoints>
This command defines the horizontal definition points of a limit line.
Suffix:
<n>
.
irrelevant
<li> Limit line
Parameters:
<LimitLinePoints> Variable number of x-axis values.
Note that the number of horizontal values has to be the same as
the number of vertical values set with CALCulate<n>:
LIMit<li>:LOWer[:DATA] or CALCulate<n>:LIMit<li>:
UPPer[:DATA]. If not, the R&S ESW either adds missing val-
ues or ignores surplus values.
*RST: -
Default unit: HZ
Manual operation: See "Data Points" on page 250
CALCulate<n>:LIMit<li>:CONTrol:DOMain <SpanSetting>
This command selects the domain of the limit line.
Suffix:
<n>
.
irrelevant
<li> Limit line
Parameters:
<SpanSetting> FREQuency | TIME
FREQuency
For limit lines that apply to a range of frequencies.
TIME
For limit lines that apply to a period of time.
*RST: FREQuency
Example:
CALC:LIM:CONT:DOM FREQ
Select a limit line in the frequency domain.
Manual operation: See "X-Axis" on page 250
CALCulate<n>:LIMit<li>:CONTrol:MODE <Mode>
This command selects the horizontal limit line scaling.