Remote Commands in the Receiver Application
R&S
®
ESW
573User Manual 1177.6298.02 ─ 08
CALCulate<n>:LIMit<li>:NAME <Name>
This command selects a limit line that already exists or defines a name for a new limit
line.
Suffix:
<n>
.
Window
<li> Limit line
Parameters:
<Name> String containing the limit line name.
*RST: REM1 to REM8 for lines 1 to 8
Manual operation: See "Name" on page 249
CALCulate<n>:LIMit<li>:UNIT <Unit>
This command defines the unit of a limit line.
Suffix:
<n>
.
irrelevant
<li> Limit line
Parameters:
<Unit> If you select dB as the limit line unit, the command automatically
turns the limit line into a relative limit line.
*RST: DBM
Manual operation: See "Y-Axis" on page 250
CALCulate<n>:LIMit<li>:UPPer[:DATA] <LimitLinePoints>
This command defines the vertical definition points of an upper limit line.
Suffix:
<n>
.
irrelevant
<li> Limit line
Parameters:
<LimitLinePoints> Variable number of level values.
Note that the number of vertical values has to be the same as
the number of horizontal values set with CALCulate<n>:
LIMit<li>:CONTrol[:DATA]. If not, the R&S ESW either
adds missing values or ignores surplus values.
*RST: Limit line state is OFF
Default unit: DBM
Manual operation: See "Data Points" on page 250