Remote Commands in the Receiver Application
R&S
®
ESW
572User Manual 1177.6298.02 ─ 08
CALCulate<n>:LIMit<li>:LOWer:SHIFt <Distance>
This command moves a complete lower limit line vertically.
Compared to defining an offset, this command actually changes the limit line definition
points by the value you define.
Suffix:
<n>
.
Window
<li> Limit line
Parameters:
<Distance> Defines the distance that the limit line moves.
Default unit: DB
Manual operation: See "Shift y" on page 251
CALCulate<n>:LIMit<li>:LOWer:SPACing
<InterpolType>
This command selects linear or logarithmic interpolation for the calculation of a lower
limit line from one horizontal point to the next.
Suffix:
<n>
.
Window
<li> Limit line
Parameters:
<InterpolType> LINear | LOGarithmic
*RST: LIN
Manual operation: See "Y-Axis" on page 250
CALCulate<n>:LIMit<li>:LOWer:THReshold <Threshold>
This command defines a threshold for relative limit lines.
The R&S ESW uses the threshold for the limit check, if the limit line violates the thresh-
old.
Suffix:
<n>
.
irrelevant
<li> Limit line
Parameters:
<Threshold> Numeric value.
The unit depends on CALCulate<n>:LIMit<li>:UNIT
on page 573.
*RST: -200 dBm
Default unit: DBM
Manual operation: See "Threshold" on page 249