Remote Commands in the Receiver Application
R&S
®
ESW
571User Manual 1177.6298.02 ─ 08
Suffix:
<n>
.
irrelevant
<li> Limit line
Parameters:
<LimitLinePoints> Variable number of level values.
Note that the number of vertical values has to be the same as
the number of horizontal values set with CALCulate<n>:
LIMit<li>:CONTrol[:DATA]. If not, the R&S ESW either
adds missing values or ignores surplus values.
*RST: Limit line state is OFF
Default unit: DBM
Manual operation: See "Data Points" on page 250
CALCulate<n>:LIMit<li>:LOWer:MARGin
<Margin>
This command defines an area around a lower limit line where limit check violations
are still tolerated.
Suffix:
<n>
.
irrelevant
<li> Limit line
Parameters:
<Margin> numeric value
*RST: 0
Default unit: dB
Manual operation: See "Margin" on page 250
CALCulate<n>:LIMit<li>:LOWer:MODE <Mode>
This command selects the vertical limit line scaling.
Suffix:
<n>
.
Window
<li> Limit line
Parameters:
<Mode> ABSolute
Limit line is defined by absolute physical values.
The unit is variable.
RELative
Limit line is defined by relative values related to the reference
level (dB).
*RST: ABSolute
Manual operation: See "X-Axis" on page 250
Analysis

Table of Contents