Remote Commands in the Receiver Application
R&S
®
ESW
570User Manual 1177.6298.02 ─ 08
Suffix:
<n>
.
irrelevant
<li> Limit line
Parameters:
<Mode> ABSolute
Limit line is defined by absolute physical values (Hz or s).
RELative
Limit line is defined by relative values related to the center fre-
quency (frequency domain) or the left diagram border (time
domain).
*RST: ABSolute
CALCulate<n>:LIMit<li>:CONTrol:SHIFt
<Distance>
This command moves a complete limit line horizontally.
Compared to defining an offset, this command actually changes the limit line definition
points by the value you define.
Suffix:
<n>
.
irrelevant
<li> Limit line
Parameters:
<Distance> Numeric value.
The unit depends on the scale of the x-axis.
Default unit: HZ
Manual operation: See "Shift x" on page 251
CALCulate<n>:LIMit<li>:CONTrol:SPACing <InterpolMode>
This command selects linear or logarithmic interpolation for the calculation of limit lines
from one horizontal point to the next.
Suffix:
<n>
.
Window
<li> Limit line
Parameters:
<InterpolMode> LINear | LOGarithmic
*RST: LIN
Example:
CALC:LIM:CONT:SPAC LIN
Manual operation: See "X-Axis" on page 250
CALCulate<n>:LIMit<li>:LOWer[:DATA] <LimitLinePoints>
This command defines the vertical definition points of a lower limit line.
Analysis

Table of Contents