Remote Commands in the Receiver Application
R&S
®
ESW
575User Manual 1177.6298.02 ─ 08
CALCulate<n>:LIMit<li>:UPPer:SPACing <InterpolType>
This command selects linear or logarithmic interpolation for the calculation of an upper
limit line from one horizontal point to the next.
Suffix:
<n>
.
Window
<li> Limit line
Parameters:
<InterpolType> LINear | LOGarithmic
*RST: LIN
Manual operation: See "Y-Axis" on page 250
CALCulate<n>:LIMit<li>:UPPer:THReshold <Limit>
This command defines an absolute limit for limit lines with a relative scale.
The R&S ESW uses the threshold for the limit check, if the limit line violates the thresh-
old.
Suffix:
<n>
.
irrelevant
<li> Limit line
Parameters:
<Limit> Numeric value.
The unit depends on CALCulate<n>:LIMit<li>:UNIT
on page 573.
*RST: -200
Default unit: dBm
Manual operation: See "Threshold" on page 249
Limit Check
CALCulate<n>:LIMit<li>:CLEar[:IMMediate]......................................................................575
CALCulate<n>:LIMit<li>:FAIL?........................................................................................ 576
CALCulate<n>:LIMit<li>:CLEar[:IMMediate]
This command deletes the result of the current limit check.
The command works on all limit lines in all measurement windows at the same time.
Suffix:
<n>
.
Window
<li> irrelevant
Example:
CALC:LIM:CLE
Deletes the result of the limit check.
Analysis