Remote Commands in the Receiver Application
R&S
®
ESW
608User Manual 1177.6298.02 ─ 08
Manual operation: See "Custom information about the measurement" on page 295
HCOPy:TREPort:ITEM:HEADer:LINE<line>:TEXT <Description>
This command defines a descriptive text for one of the items part of the report header.
You can define up to 6 items in the header.
Use HCOPy:TREPort:ITEM:HEADer:LINE<line>:TITLe on page 608 to define
custom titles for each item.
Use HCOPy:TREPort:ITEM:HEADer:LINE<line>:CONTrol to select the condition
under which each item is shown.
Suffix:
<line>
.
1...6
Selects the header line.
Parameters:
<Description> String containing the description of one of the value fields.
By default, the value fields of the items are empty.
Example:
HCOP:TREP:ITEM:HEAD:LINE3:TITL 'Device under
Test'
Renames the third title into "Device under Test".
HCOP:TREP:ITEM:HEAD:LINE3:TEXT 'Some Device'
Labels the third title as "Some Device".
Manual operation: See "Custom information about the measurement" on page 295
HCOPy:TREPort:ITEM:HEADer:LINE<line>:TITLe <Title>
This command defines a custom name for one of the items part of the report header.
You can define up to 6 items in the header.
Use HCOPy:TREPort:ITEM:HEADer:LINE<line>:TEXT to add a value to each
item.
Use HCOPy:TREPort:ITEM:HEADer:LINE<line>:CONTrol to select the condition
under which each item is shown.
Suffix:
<line>
.
1...6
Selects the header line.
Parameters:
<Title> String containing the title of the item.
The default titles are as follows:
• Line 1: "Heading"
• Line 2: "Meas Type"
• Line 3: "Equipment under Test"
• Line 4: "Manufacturer"
• Line 5: "OP Condition"
• Line 6: "Test Spec"
Data Management