Remote Commands in the Receiver Application
R&S
®
ESW
611User Manual 1177.6298.02 ─ 08
Note that the items, separated by commas, have to be written
into one string (see example below).
The available items depend on the application you are using.
See the tables below for a short description of each item.
Per default, some items are selected (see tables below).
Example:
HCOP:TREP:ITEM:SEL 'SETT,MARK,SRES,DIAG'
Each dataset consists of the measurement settings, marker
information, the scan results and a screenshot of the scan trace
(for Receiver channels).
The selection is applied to the currently selected channel.
Example:
HCOP:TREP:ITEM:SEL REC,'BARG,LISN'
A dataset in the Receiver application consists of the bargraph
and the LISN settings.
Manual operation: See "Test report content selection" on page 293
Table 15-6: Available <ChannelTypes>
<ChannelType> Description
ADEM Analog demodulator
CAPD CISPR APD
IQ I/Q analyzer
REC Receiver
RTIM Real-time
SAN Spectrum
Table 15-7: Available <items> in receiver application
<Item> Description Default
BARGraph Screenshot of the bargraph.
DIAGram Screenshot of the scan results. x
FRESults Numerical results for the final measurement.
IF Screenshot of the results for IF analysis. x
IFSPectrogram Screenshot of the IF spectrogram. x
LISN Information about LISNs.
MARKers Contents of the marker table. x
PRESults Contents of the peak list.
SCANtable Measurement configuration as defined in the scan table.
SETTings Settings that have been used during a measurement. x
SPECtrogram Screenshot of the spectrogram. x
SRESults Numerical results for the scan.
TRANsducer Characteristics of the transducer.
Data Management