Remote Commands in the Receiver Application
R&S
®
ESW
634User Manual 1177.6298.02 ─ 08
Parameters:
<Commands> String of remote commands.
If you send more than one command, separate them with a sem-
icolon.
For a comprehensive description of commands supported by the
switch matrix, refer to its documentation.
Example:
CORR:SWIT:RANG4 'ROUT:CLOS (@F01A11(0101))'
[SENSe:]CORRection:SWITch:RANGe<range>:EXECute
This command deliberately sends the commands defined for a certain transducer
range to the switch matrix.
Suffix:
<range>
.
1...10
Selects the transducer range.
Example:
CORR:SWIT:RANG4:EXEC
Sends the commands defined for transducer range 4.
Usage: Event
[SENSe:]CORRection:SWITch:SCPI <State>
This command turns the use of remote commands that comply to the SCPI standard
on and off.
Required for switch matrixes that support a command set that does not comply to the
SCPI standard.
Parameters:
<State> ON | OFF
*RST: ON
Example:
CORR:SWIT:SCPI ON
Command syntax has to comply to the SCPI standard.
Manual operation: See "RF switch control" on page 325
[SENSe:]CORRection:SWITch:SELect <FileName>
This command selects a dataset that controls a switch matrix.
Note that you have to select a dataset before you can edit it.
Parameters:
<FileName> String containing the file name and path.
Example:
CORR:SWIT:SEL 'dataset'
Selects a dataset called "dataset".
Manual operation: See "Dataset name and file name" on page 325