Remote Commands in the Receiver Application
R&S
®
ESW
662User Manual 1177.6298.02 ─ 08
You can define the message with SYSTem:DISPlay:MESSage[:TEXT].
Parameters:
<State> ON
Displays a custom message.
Shows a message "Remote Mode controlled by <string>",
including the network address of the controlling device.
OFF
Displays the standard message ("Remote mode").
*RST: OFF
Example: //Define and display custom message
SYST:DISP:MESS 'a computer'
SYST:DISP:MESS:STAT ON
//would display:
Remote mode controlled by a computer
SYSTem:DISPlay:UPDate <State>
This command turns the display during remote operation on and off.
If on, the R&S ESW updates the diagrams, traces and display fields only.
The best performance is obtained if the display is off during remote control operation.
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Example:
SYST:DISP:UPD ON
Manual operation: See "Remote Display Update" on page 401
SYSTem:ERRor:CLEar:REMote
This command deletes all contents of the "Remote Errors" table.
Note: The remote error list is automatically cleared when the R&S ESW is shut down.
Example:
SYST:ERR:CLE:REM
Manual operation: See "Display Remote Errors" on page 402
See "Clear Error List" on page 408
SYSTem:ERRor:DISPlay <State>
This command the error display during remote operation on and off.
If activated, the R&S ESW displays a message box at the bottom of the screen that
contains the most recent type of error and the command that caused the error.
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
General Instrument Setup