Remote Control Commands
R&S
®
SGS100A
255User Manual 1173.9105.02 ─ 12
SOURCe:PULM:TRAin:SEL ‘P_FIVE'
// Enter the pulse train data
SOURCe:PULM:TRAin:ONTime 10ns,30ns,40ns,20ns,10ns
SOURCe:PULM:TRAin:OFFTime 30ns,40ns,50ns,40ns,30ns
SOURCe:PULM:TRAin:REPetition 10,1,3,10,6
// ******************************************************************
// Select pulse train mode
// ******************************************************************
// Select the internal modulation generator and the pulse mode
SOURce:PULM:SOURce INTernal
SOURce:PULM:MODE PTRain
// ******************************************************************
// Activate the signal output
// ******************************************************************
SOURce:PGENerator:OUTPut:STATe 1
SOURce:PULM:STATe 1
OUTPut1:STATe 1
[:SOURce<hw>]:PULM:DELay........................................................................................ 255
[:SOURce<hw>]:PULM:DOUBle:DELay............................................................................256
[:SOURce<hw>]:PULM:DOUBle:WIDTh........................................................................... 256
[:SOURce<hw>]:PULM:MODE........................................................................................ 256
[:SOURce<hw>]:PULM:PERiod....................................................................................... 256
[:SOURce<hw>]:PULM:POLarity..................................................................................... 257
[:SOURce<hw>]:PULM:SOURce..................................................................................... 257
[:SOURce<hw>]:PULM:STATe.........................................................................................257
[:SOURce<hw>]:PULM:TRIGger:EXTernal:GATE:POLarity.................................................258
[:SOURce<hw>]:PULM:TRIGger:EXTernal:IMPedance...................................................... 258
[:SOURce<hw>]:PULM:TRIGger:EXTernal:SLOPe............................................................ 258
[:SOURce<hw>]:PULM:TRIGger:MODE...........................................................................258
[:SOURce<hw>]:PULM:WIDTh........................................................................................259
[:SOURce<hw>]:PULM:DELay <Delay>
Sets the pulse delay.
Parameters:
<Delay> float
Range: 0 to 100 s
Increment: 10 ns
*RST: 10 ns
Example:
PULM:DEL 13 us
13 us elapse after a trigger before the first pulse is generated.
Manual operation: See "Pulse Delay" on page 116