List and Sweep Mode
R&S
®
SMA100B
164User Manual 1178.3834.02 ─ 05
Configuration and operation of sweep mode signals
●
The R&S SMA100B generates a sweep signal by varying one of the following
parameters: the RF frequency, the LF frequency or the RF level.
●
In all sweep modes, you can perform a complete sweep cycle once, repeat the
cycle continuously or step through it gradually.
●
The instrument generates the sweep signal according to the sweep settings.
●
The "Reset Sweep" function, provided in the sweep settings dialogs enables you to
reset the sweep to its initial value and restart it.
Impact of changing the sweep mode during performance
If you change the sweep mode during the execution, the signal generator stops the
sweep and starts with the next trigger event at the initial value.
6.2.1 Correlating Parameters in Sweep Mode
A sweep signal is a periodic signal that changes its frequency or level from a starting
value to an ending value in a defined time.
The R&S SMA100B provides various possibilities to configure a sweep signal. For
example, you can determine the sweep range by the start and end values, or based on
the start value and span. In any case, related parameters are adjusted accordingly.
The formulas below show how the sweep parameters correlate and the corresponding
calculation basis, by the frequency and offset settings. Apart from "Center Frequency",
"Span" and "Step_lin", the values apply accordingly to the level settings.
Table 6-6: Variables that are used in the following formulas
Variable Description
Sweep range Defined frequency or level value range
f
CENTer
Defined center frequency
f
SPAN
Defined extend of the sweep range
f
OFFSet
Frequency offset
f
STARt
Start frequency of the sweep range
f
STOP
End frequency of the sweep range
f
1
Current sweep frequency
f
2
Next, subsequent sweep frequency
step_lin Step size in linear scaling
step_log Step size in logarithmic scaling
POINts Number of steps within the sweep range
About Sweep Mode