Improving Level Performance
R&S
®
SMA100B
263User Manual 1178.3834.02 ─ 05
10. Close the "Configure" dialog.
11. Press "Start Cont" to start the measurement.
7.4.4.11 How to Setup a Pulse Measurement
Prerequisites: the sensor is connected to the instrument and configured, see Chap-
ter 7.4, "Using Power Sensors", on page 211.
1. Select "Clk Syn/Power Sens" > "NRP Power Analysis".
2. Select "Configure" > "Configure Measurement".
3. Set "Mode" > "Time".
4. Configure the measurement:
a) Set "Min = -5 us".
b) Set "Max = 45 us".
c) Set "Steps = 200".
5. Select the "Configure Diagram" tab. Configure the diagram, see Chapter 7.4.4.12,
"How to Configure the Power Analysis Diagram", on page 264.
6. Select the "Configure Sensors" tab.
7. Select the connected sensor, e.g. "2-NRP-Z81".
8. In the "Next Window List" tab, enable the "Pulse Data View".
9. Close the "Configure" dialog.
10. In the diagram window, open the context-sensitive menu.
11. Select "Trace/Pulse Data Notifications/ Trigger".
12. Select the "Trace" tab.
13. Set "State > On".
14. Set the thresholds:
a) Set "Base > Voltage Related".
b) Set "Distal = 90%".
c) "Mesial = 50%".
d) "Proximal = 10%".
15. Select the "Pulse Data Notifications" tab.
16. Enable up to six parameters, for example:
"Duty Cycle"
"Pulse Width"
"Pulse Period"
"Off Time"
"Start Time"