List and Sweep Mode
R&S
®
SMA100B
175User Manual 1178.3834.02 ─ 05
Note: The minimum dwell time varies, depending on the phase noise options (Option:
R&S SMAB-B709, R&S SMAB-B710/-B710N or R&S SMAB-B711/-B711N) the instru-
ment is equipped with. See the data sheet for the specifications.
In case of considerable overrun conditions, the R&S SMA100B turns off the sweep
mode.
See also Chapter 6.4, "Significant Parameters and Functions", on page 168.
Remote command:
[:SOURce<hw>]:SWEep[:FREQuency]:DWELl on page 648
[:SOURce<hw>]:SWEep:POWer:DWELl on page 645
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:DWELl on page 609
Trigger Slope
For "Mode = Extern Step/Single", selects the polarity of the active slope of an applied
instrument trigger.
Trigger signal is expected at the Inst Trig connector.
"Positive"
The rising edge of the trigger signal triggers the instrument.
"Negative"
The falling edge of the trigger signal triggers the instrument.
Remote command:
[:SOURce]:INPut:TRIGger:SLOPe on page 599
Reset Sweep
Resets a sweep.
With the next trigger event, the sweep starts at the initial value.
Remote command:
[:SOURce<hw>]:SWEep:RESet[:ALL] on page 651
Execute Single Sweep
In "Mode = Single", starts a sweep manually.
Remote command:
[:SOURce<hw>]:SWEep[:FREQuency]:EXECute on page 650
[:SOURce<hw>]:SWEep:POWer:EXECute on page 650
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:EXECute on page 609
:TRIGger<hw>[:SWEep][:IMMediate] on page 684
:TRIGger<hw>:FSWeep[:IMMediate] on page 684
:TRIGger<hw>:PSWeep[:IMMediate] on page 684
:TRIGger<hw>:LFFSweep:IMMediate on page 684
6.5.2 Frequency Range Settings
Access:
1. Select for example "Sweep" > "RF Frequency Sweep"
2. Select "Frequency Range".
Sweep Mode Settings