Reference Oscillator
R&S
®
SMA100B
276User Manual 1178.3834.02 ─ 05
"1 GHz"
Outputs a 1 GHz signal.
"Input Signal (loop through)"
Forwards the external reference frequency to the output directly.
Remote command:
[:SOURce]:ROSCillator:OUTPut:FREQuency:MODE on page 641
[:SOURce]:ROSCillator:OUTPut:ALTernate:FREQuency:MODE on page 642
Show Connector
Accesses a dialog that displays the physical location of the selected connector on the
front/rear panel of the instrument.
8.5 Adjustment Settings
Access:
1. Select "Frequency" > "Reference Freq"
2. Select "Adjustment".
Settings:
Adjustment Active....................................................................................................... 276
Adjustment Value........................................................................................................ 276
Adjustment Active
Selects the adjustment mode.
"OFF"
Uses the calibrated internal reference frequency.
"ON"
Allows you to apply a deviation to the internal reference frequency,
according to your requirements.
Enter the value in the Adjustment Value field.
Remote command:
[:SOURce]:ROSCillator[:INTernal]:ADJust[:STATe] on page 643
Adjustment Value
Sets a user-defined adjustment value for the internal reference frequency. This value
takes effect when it is activated with Adjustment Active.
"0" represents the calibrated state.
The setting range depends on the reference oscillator type and its factory calibra-
tion value.
Note:
The setting is not affected by an instrument preset ([Preset] key or *RST) and the
"Save/Recall" function. It is reset only by factory preset.
Adjustment Settings