Troubleshooting and Error Messages
R&S
®
SMA100B
704User Manual 1178.3834.02 ─ 05
Error
Code
Error Description Remedy
183 Adjustment data invalid Adjustment data is invalid and must be restored. Generate the adjustment data and load it
into the device
200 Cannot access hardware The data transmission to a module was unsuc-
cessful.
The module is not installed, not properly
installed or missing.
201 Hardware revision out of
date
A later version of certain parts of the instrument is
necessary to execute the function selected.
The driver does not support the installed
version of a module.
202 Cannot access the
EEPROM
An error occurs when writing or reading a
EEPROM.
The EEPROM is possibly defect.
Replace it.
203 Invalid EEPROM data Reading a EEPROM is possible, however the data
are inconsistent.
204 Driver initialization failed Initialization of a driver fails when booting the
instrument firmware.
The driver is not compatible with the
hardware or software configuration of the
instrument.
241 No current list There is no list selected. To execute the required
operation, a list has to be selected in the related
dialog. If no list is available, a new list must be cre-
ated.
242 Unknown list type speci-
fied
The list type selected is not valid for the required
operation.
Check the selected list type.
460 Cannot open the file The selected file cannot be opened. Check the path and file name.
461 Cannot write file The file cannot be written. Check if the file is read-only.
462 Cannot read file The file cannot be read. Check if the file contents are compatible
with the file type.
463 Filename missing The required operation cannot be executed
because the file name is not specified.
Enter file name when creating list.
464 Invalid filename extension The file extension is not valid for the required
operation.
Check the file extension.
465 File contains invalid data The selected file contains data that is not valid for
the file type.
The file extension determines the data that is valid
for this file type. If the file extension is changed,
the lists are no longer recognized and the data is
therefore invalid.
Check the file extension.
15.4 Querying Error Messages
The R&S SMA100B monitors the functions performed and automatically detects errors
and irregularities. The instrument displays the messages in the "Info" line and records
them in the background with a detailed description.
Find details to the system messages in Chapter 15, "Troubleshooting and Error Mes-
sages", on page 702.
Querying Error Messages