File and Data Management
R&S
®
SMA100B
294User Manual 1178.3834.02 ─ 05
Remote command:
[:SOURce<hw>]:FREQuency[:CW|FIXed]:RCL on page 593
Exclude Level
The current level is retained when a stored instrument setting is loaded.
Remote command:
[:SOURce<hw>]:POWer[:LEVel][:IMMediate]:RCL on page 635
Recall
Restores the selected configuration.
During recall, the instrument considers all related settings, for example sweeps in
active state or lists. An error message indicates the settings which cannot be imple-
mented.
Remote command:
:MMEMory:LOAD:STATe on page 420
Recall Immediate x
Loads the selected configuration from one of the intermediate memories. A message
appears if no instrument configuration is stored in this memory.
Remote command:
*RCL on page 409
File Manager
Accesses the "File Manager" dialog, see Chapter 10.8, "Using the File Manager",
on page 298.
10.4.2 How to Save and Recall Instrument Settings
Instrument settings can be saved to a file and loaded again later, so that you can
repeat the tests with the same settings.
Accessing and recalling instrument setups
For quick access to a stored instrument setup, assign the appropriate action to the [
(User)] key.
See Chapter 11.2.3, "Assigning Actions to the [ (User)] Key", on page 326.
To save and recall instrument settings quickly
1. Select "System Config > Save/Recall" > "Operation Mode > Save".
2. Select "Save Immediate 1".
The instrument saves its settings in the intermediate memory 1. The filename and
the storage location cannot be changed.
3. Adapt the instrument settings as required. Select "Save Immediate 2"
4. To restore the settings, select the "Operation Mode > Recall"
Saving and Recalling Instrument Settings