Improving Level Performance
R&S
®
SMA100B
207User Manual 1178.3834.02 ─ 05
7.3.3 Fill with Sensor
Access:
1. Select "Level" > "User Correction".
2. Select "UCOR Data > navigate to the file *.ucor > Select".
3. Select "Edit UCOR Data > Fill With Sensor...".
This dialog contains parameters for filling a table automatically with sensor read-
ings.
To fill the table, select "Execute".
For information on power sensors an how to use them, see Chapter 7.5, "How to Cali-
brate the Power Level with an R&S NRP Power Sensor", on page 264.
Settings
Fill User Correction Data with Sensor......................................................................... 207
Used SMAB Settings For Measurement..................................................................... 208
Fill User Correction Data with Sensor
"Sensor"
Displays connected sensors for selection.
"List To Fill"
Indicates the used list.
"Use SParameter"
Indicates whether SParameter correction in the R&S NRP power sensor is used.
"Include Zeroing"
Performs a zeroing procedure before acquiring the user correction data to improve
precision.
No signal is applied to the sensor during zeroing. RF output is temporarily switched
off during that time.