Prevents servicing of the subsequent commands until all preceding commands have
been executed and all signals have settled (see also command synchronization and
*OPC).
Usage: Event
13.4 Preset Commands
The preset commands are not bundled in one subsystem. Therefore, they are listed
separately in this section.
Four presetting actions are available:
●
Activating the default state of all internal instrument functions (*RST on page 409).
Functions that concern the integration of the instrument into a measurement setup
are not changed, e.g. GPIB address or reference oscillator source settings.
●
Activating the preset state of the parameters related to the selected signal path (:
SOURce<hw>:PRESet on page 411)
●
Activating the preset state of all parameters that are not related to the signal path
(:DEVice:PRESet on page 411)
●
Activating the original state of delivery (factory reset, :SYSTem:FPReset
on page 412). Only functions that are protected by a password remain unchanged
as well as the passwords themselves.
:DEVice:PRESet............................................................................................................ 411
:SOURce<hw>:PRESet.................................................................................................. 411
:SYSTem:PRESet.......................................................................................................... 412
:SYSTem:FPReset......................................................................................................... 412
:DEVice:PRESet
Presets all parameters which are not related to the signal path, including the LF gener-
ator.
Example:
DEV:PRES
Presets all instruments settings that are not related to the signal
path.
Usage: Event
:SOURce<hw>:PRESet
Presets all parameters which are related to the selected signal path.
Example:
SOUR:PRES
Presets all settings that are related to signal path
Usage: Event