Improving Level Performance
R&S
®
SMA100B
234User Manual 1178.3834.02 ─ 05
"Logarithmic"
Power versus frequency
In a logarithmic sweep, the frequency is swept in equidistant steps on
a logarithmic scale. The x-axis is a logarithmic frequency axis.
Remote command:
:SENSe[:POWer]:SWEep:FREQuency:SPACing[:MODE] on page 467
:SENSe[:POWer]:SWEep:POWer:SPACing[:MODE] on page 480
:SENSe[:POWer]:SWEep:TIME:SPACing[:MODE] on page 485
Execution - Power Analysis
Selects single or continuous mode in power analysis.
The measurement is started in the diagram using the "Start" button. During measure-
ment, the "Start" button is replaced by a "Stop" button which can be used to abort the
measurement. The progress bar indicates the status of the measurement.
"Single"
Selects single measurement.
"Cont."
Selects continuous measurements.
Remote command:
:SENSe[:POWer]:SWEep:RMODe on page 483
:SENSe[:POWer]:SWEep:POWer:RMODe on page 479
:SENSe[:POWer]:SWEep:FREQuency:RMODe on page 466
:SENSe[:POWer]:SWEep:TIME:RMODe on page 484
All Trigger Events - Power Analysis
Determines, whether the measurement data processing starts with a trigger event in
one of the sensors (Logical OR), or whether all channels have to be triggered (logical
AND). Each sensor evaluates a trigger event according to its setting independently.
This function supports the internal or external trigger modes with multi-channel time
measurements.
"Wired AND"
the measurement starts when all channels are triggered.
"Wired OR"
The measurement starts when a trigger event occurs.
Remote command:
:SENSe[:POWer]:SWEep:TIME:TEVents on page 486
Configure Diagram
Access:
1. Select "Clk Syn/Power Sens" > "NRP Power Analysis".
2. Select "Configure" > "Configure Diagram"
Using Power Sensors