Remote Control Commands
R&S
®
SMA100B
626User Manual 1178.3834.02 ─ 05
[:SOURce<hw>]:NOISe:DISTribution <Distribution>
Sets the distribution of the noise power density.
Parameters:
<Distribution> GAUSs | EQUal
*RST: GAUSs
Example: See Example "Configuring the noise generator" on page 625.
Manual operation: See "Distribution" on page 103
[:SOURce<hw>]:NOISe:LEVel:RELative?
Queries the level of the noise signal per Hz in the total bandwidth.
Return values:
<Relative> float
Range: -149.18 to -52.67
Increment: 0.1
*RST: -69.84
Example: See Example "Configuring the noise generator" on page 625.
Usage: Query only
Manual operation: See "Noise Density" on page 105
[:SOURce<hw>]:NOISe:LEVel[:ABSolute]?
Queries the level of the noise signal in the system bandwidth within the enabled band-
width limitation.
Return values:
<Absolute> float
Noise level within the bandwidth limitation
*RST: 3.84 MHz
Example: See Example "Configuring the noise generator" on page 625.
Usage: Query only
Manual operation: See "Noise Level" on page 106
13.16.9 SOURce:PGEN Subsystem
The PGEN subsystem contains the commands for setting output of the pulse modula-
tion signal.
SOURce Subsystem