File and Data Management
R&S
®
SMA100B
290User Manual 1178.3834.02 ─ 05
Parameter Preset value Preset Factory
Preset
Display and keyboard settings - - x
Password and settings protected by passwords (e.g. dis-
abled LAN or USB)
- - -
Security settings - - -
User files (setups
2)
, data lists, etc.)
- - -
●
2)
UserPreset.savrcltxt is renamed as UserPresetInactive.savrcltxt;
an existing file with the same name is overwritten.
If the default values in the "Remote Access" dialog had been changed, a factory preset
via remote control (:SYSTem:FPReset) terminates the connection to the instrument.
Security settings are never reset.
Resets all parameters and switching states, and closes all opened dialogs.
10.3 Protecting Data
During operation, the R&S SMA100B saves user data permanently in the user direc-
tory, see "File storage location" on page 283.
To protect any classified data and to avoid saving any sensitive data on the
R&S SMA100B permanently, you can:
●
Install the option removable memory R&S SMAB-B85
This option ensures that user data is never saved on the internal memory.
– Per default, if removable memory is mounted, user data is saved perma-
nently and only on this storage medium.
You can access data saved on the removable memory just as data stored in
the /var/user/.
– If volatile mode is activated or there is no memory mounted or it is removed
during operation, user data is stored temporarily in the volatile memory of
the instrument.
This data is lost once the instrument is switched off.
●
Store user files temporarily in the /var/volatile directory, which remains
available only until the instrument is switched off.
You can access data saved in volatile memory just as data stored permanently in
the /var/user/.
See also Chapter 10.8.3, "How to Display All Saved Files", on page 302.
●
Activate the volatile mode so that no user data can be written to the internal mem-
ory permanently.
The removable memory is protected, too.
Protecting Data