Network Operation and Remote Control
R&S
®
SMA100B
355User Manual 1178.3834.02 ─ 05
12.3.2 Possible Setups and Access Functions
The following examples give an overview of dependencies between the available libra-
ries, the possible interfaces and protocols, and whether an instrument driver is provi-
ded. The involved parts are highlighted. For more information, see the application
note 1GP72: Connectivity of Rohde&Schwarz Signal Generators.
Remote control (application) program using VISA
Figure 12-3: Application program using VISA
Protocol Remote control program
Socket
viOpen (..., "TCPIP:SMA100B-102030::5025::SOCKET", ...)
viPrintf (..., "SOUR:FREQ 2GHz\n")
VXI-11
viOpen (..., "TCPIP:SMA100B-102030::inst0::INSTR", ...)
viPrintf (..., "SOUR:FREQ 2GHz\n")
HiSLIP
viOpen (..., "TCPIP:SMA100B-102030::hislip0::INSTR", ...)
viPrintf (..., "SOUR:FREQ 2GHz\n")
USBTMC
viOpen (..., "USB::0x0AAD::0x01DD::100001::INSTR", ...)
viPrintf (..., "SOUR:FREQ 2GHz\n")
GPIB
viOpen (..., "GPIB::28::INSTR", ...)
viPrintf (..., "SOUR:FREQ 2GHz\n")