List and Sweep Mode
R&S
®
SMA100B
177User Manual 1178.3834.02 ─ 05
"Linear"
Takes the frequency value entered as absolute value in Hz.
"Logarithmic"
Takes the value entered as a logarithmic value, i.e. as a constant
fraction of the current frequency in %.
Remote command:
[:SOURce<hw>]:SWEep[:FREQuency]:SPACing on page 649
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:SPACing on page 611
Step Linear/Step Logarithmic
In "RF/LF Frequency Sweep" mode, sets the step width for the individual frequency
sweep steps. The value is added at each sweep step to the current frequency.
Depending on the current Spacing, you can enter either an absolute or logarithmic step
width.
This parameter is only available for Sweep Type > Step.
See Chapter 6.2.1, "Correlating Parameters in Sweep Mode", on page 164.
"Step Linear"
The step width is a constant value in Hz.
"Step Logarithmic"
The step width is determined logarithmically in %, i.e. as a constant
fraction of the current frequency.
Remote command:
[:SOURce<hw>]:SWEep[:FREQuency]:STEP[:LINear] on page 651
[:SOURce<hw>]:SWEep[:FREQuency]:STEP:LOGarithmic on page 650
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:STEP[:LINear] on page 612
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:STEP:LOGarithmic
on page 611
6.5.3 Level Range Settings
Access:
1. Select "Sweep" > "RF Level Sweep"
2. Select "Level Range".
Sweep Mode Settings