585User Manual 1178.3834.02 ─ 05
// Select source and destination
SOURce1:CORRection:DEXChange:AFILe:SELect "/var/user/ucor2ASCII"
// Export the user correction data into the ASCII file
SOURce1:CORRection:DEXChange:MODE EXPort
SOURce1:CORRection:DEXChange:EXECute
// Query the available ASCII files with extension .csv
SOURce1:CORRection:DEXChange:AFILe:CATalog?
// ucor2ASCII,my_ucor
// Import a user correction ASCII file
SOURce1:CORRection:DEXChange:MODE IMPort
SOURce1:CORRection:DEXChange:AFILE:SELect "/var/user/my_ucor"
SOURce1:CORRection:DEXChange:EXECute
Correction Settings................................................................................................585
Correction Data Exchange....................................................................................588
13.16.3.1 Correction Settings
[:SOURce<hw>]:CORRection:CSET:DATA:FREQuency <Frequency>
Enters the frequency value in the table selected with [:SOURce<hw>]:CORRection:
CSET[:SELect].
Parameters:
<Frequency> Frequency#1[, Frequency#2, ...]
String of values with default unit Hz.
Example: See Example "Create a table with user-defined correction values
for the RF level" on page 583 .
Manual operation: See "Edit List Mode Data" on page 190
[:SOURce<hw>]:CORRection:CSET:DATA:POWer <Power>
Enters the level values to the table selected with [:SOURce<hw>]:CORRection:
CSET[:SELect].
Parameters:
<Power> Power#1[, Power#2, ...]
String of values with default unit dB.
*RST: 0
Example: See Example "Create a table with user-defined correction values
for the RF level" on page 583 .
Manual operation: See "Edit List Mode Data" on page 190