Improving Level Performance
R&S
®
SMA100B
262User Manual 1178.3834.02 ─ 05
3. Set "Mode" > "Frequency".
4. Configure the measurement:
a) Set "Min = 8KHz".
b) Set "Max = 6Ghz".
c) Set "Steps = 200".
5. Select the "Configure Diagram" tab. Configure the diagram, see Chapter 7.4.4.12,
"How to Configure the Power Analysis Diagram", on page 264.
6. Select the "Configure Sensors" tab.
7. Select the connected sensor, e.g. "2-NRP-Z81".
8. If you want that the frequency settings of the sensor are configured differently than
the generator settings, enable "Use Separate Frequ.".
a) Set "Min = 1MHz".
b) Set "Max =10Mhz".
9. In the "Next Window List" tab, enable the "Standard View".
10. Close the "Configure" dialog.
11. Press "Start Cont" to start the measurement.
7.4.4.10 How to Setup a Power Sweep Measurement
Prerequisites: the sensor is connected to the instrument and configured, see Chap-
ter 7.4, "Using Power Sensors", on page 211.
1. Select "Clk Syn/Power Sens" > "NRP Power Analysis".
2. Select "Configure" > "Configure Measurement".
3. Set "Mode" > "Power".
4. Configure the measurement:
a) Set "Min = -40 dBm".
b) Set "Max = 40dBm".
c) Set "Steps = 200".
5. Select the "Configure Diagram" tab. Configure the diagram, see Chapter 7.4.4.12,
"How to Configure the Power Analysis Diagram", on page 264.
6. Select the "Configure Sensors" tab.
7. Select the connected sensor, e.g. "2-NRP-Z81".
8. If you want that the frequency settings of the sensor are configured differently than
the generator settings, enable "Use Separate Frequ.".
a) Set "1MHz".
9. In the "Next Window List" tab, enable the "Standard View".
Using Power Sensors