Improving Level Performance
R&S
®
SMA100B
261User Manual 1178.3834.02 ─ 05
7.
In the "Automatic Naming" tab:
a) To change the default directory the file is saved in, select "Path" and define a
path and a filename. For example, select the default directory /var/user.
b) If necessary, disable or change some of the parameters in the "Automatic Nam-
ing Settings".
c) Close the "Save NRP-Z Analysis Options" dialog.
8. In the "Save NRP-Z Analysis" dialog, select "Save".
The instrument saves a hardcopy of the current instrument display as a *.jpg file.
The filename is automatically created.
9. To print the hardcopy, connect the instrument to a LAN and:
a) Transfer the file to a remote computer as described in Chapter 10.9, "How to
Transfer Files from and to the Instrument", on page 305.
b) On the remote computer, navigate through the file system.
c) Print the selected file.
For more information, refer to the online help of the operating system.
7.4.4.9 How to Setup a Frequency Sweep Measurement
Prerequisites: the sensor is connected to the instrument and configured, see Chap-
ter 7.4, "Using Power Sensors", on page 211.
1. Select "Clk Syn/Power Sens" > "NRP Power Analysis".
2. Select "Configure" > "Configure Measurement".
Using Power Sensors