Remote Control Commands
R&S
®
SMA100B
454User Manual 1178.3834.02 ─ 05
:DISPlay[:WINDow][:POWer]:SWEep:GRID:STATe <State>
Indicates a grid in the diagram.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 1
Example:
DISP:SWE:GRID:STAT OFF
Deactivates the indication of a grid in the diagram area.
Manual operation: See " Show Grid - Power Analysis " on page 236
13.14.3
INITiate Command
:INITiate<hw>[:POWer]:CONTinuous <Continuous>
The command switches the local state of the continuous power measurement by the
R&S NRP-Zxx power sensors on and off. Switching off the local state enhances the
measurement performance during remote control
The remote measurement is triggered by the READ query (command :READ<ch>[:
POWer]? on page 454) which also provides the measurement results. The local state
is not influenced by this command, measurements results can be retrieved with local
state on or off.
Parameters:
<Continuous> 0 | 1 | OFF | ON
*RST: OFF
Example:
INIT:CONT ON
switches local state of continuous power measurement on.
13.14.4 READ Subsystem
:READ<ch>[:POWer]?
The command triggers the measurement with power sensors and provides the power
measurement result of the selected power sensor. The value is provided with the unit
set with command SENSe:UNIT[:POWer].
For certain power sensors, e.g. R&S NRP-Z81, two values are returned, first the value
for the average level and - separated by a comma - the peak level
Note: The local state is not influenced by this command, measurements results can be
retrieved with local state on or off. For long measurement times it is recommended to
use a SRQ (MAV bit) for command synchronization.
Suffix:
<ch>
.
1..3
Power Sensor Measurement Subsystems