Remote Control Commands
R&S
®
SMA100B
485User Manual 1178.3834.02 ─ 05
Manual operation: See " Execution - Power Analysis " on page 234
:SENSe[:POWer]:SWEep:TIME:SPACing[:MODE] <Mode>
Queries the sweep spacing for the power versus time measurement. The spacing is
fixed to linear.
Parameters:
<Mode> LINear
*RST: LINear
Example:
SENS:SWE:TIME:SPAC?
Queries the sweep spacing.
Manual operation: See " Spacing - Power Analysis " on page 233
:SENSe[:POWer]:SWEep:TIME:STARt <Start>
Sets the start time for the power versus time measurement. Value 0 defines the trigger
point. By choosing a negative time value, the trace can be shifted in the diagram. It is
possible, that the measurement cannot be performed over the complete time range
because of limitations due to sensor settings. In this case, an error message is output.
Parameters:
<Start> float
Range: -1 to 1
Increment: 1E-12
*RST: -5E-6
Example:
SENS:SWE:TIME:STAR 0s
Sets the start time to 0 s.
Manual operation: See "Min - Power Analysis" on page 231
:SENSe[:POWer]:SWEep:TIME:STEPs <Steps>
Sets the number of measurement steps for the power versus time measurement. Value
0 defines the trigger point.
Parameters:
<Steps> integer
Range: 10 to 1000
*RST: 500
Example:
SENS:SWE:TIME:STEP 500
Sets the 500 measurement steps.
Manual operation: See " Steps - Power Analysis " on page 232
:SENSe[:POWer]:SWEep:TIME:STOP <Stop>
Sets the stop time for the power versus time measurement.
Power Sensor Measurement Subsystems