EasyManua.ls Logo

Rohde & Schwarz SMAB-B92 - Page 488

Rohde & Schwarz SMAB-B92
778 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMA100B
486User Manual 1178.3834.02 ─ 05
Parameters:
<Stop> float
Range: 0 to 2
Increment: 1E-12
*RST: 1E-3
Example:
SENS:SWE:TIME:STOP 1ms
Sets the stop time to 1 ms.
Manual operation: See "Max - Power Analysis" on page 232
:SENSe[:POWer]:SWEep:TIME:TEVents
<TriggerTEvents>
Determines, whether the measurement data processing starts with a trigger event in
one of the sensors (Logical OR), or whether all channels have to be triggered (logical
AND). Each sensor evaluates a trigger event according to its setting independently.
This function supports the internal or external trigger modes with multi-channel time
measurements.
Parameters:
<TriggerTEvents> AND | OR
*RST: AND
Example:
SENS:POW:SWE:TIME:TEV AND
The measurement data processing starts when all channels are
triggered.
Options: Option R&S SMAB-K28
Manual operation: See " All Trigger Events - Power Analysis " on page 234
:SENSe[:POWer]:SWEep:TIME:YSCale:AUTO <Auto>
Activates autoscaling of the Y axis in the diagram.
Parameters:
<Auto> OFF | CEXPanding | FEXPanding | CFLoating | FFLoating
OFF
Auto scaling is deactivated. When switching from activated to
deactivated Auto scaling, the scaling is maintained. When
switching from deactivated to activated Auto scaling, the scaling
is reset to min = max = 0.
CEXPanding | FEXPanding
Auto scale is activated. The scaling of the Y-axis is selected in
such a way, that the trace is always visible. To this end, the
range is expanded if the minimum or maximum values of the
trace move outside the current scale. The step width is 5 dB for
selection course and variable in the range of 0.2 db to 5 dB for
selection fine.
Power Sensor Measurement Subsystems

Table of Contents