Remote Control Commands
R&S
®
SMA100B
452User Manual 1178.3834.02 ─ 05
Example:
SENS:SWE:MODE TIME
Activates time mode for power analysis.
CALC:SWE:TIME:GATE:STAT ON
Activates time gated measurement.
SENS:SWE:INIT
Activates a single power analysis measurement.
CALC:SWE:TIME:GATE2:MAX?
Queries the peak power in time gate 2 for trace 1 (=default).
Usage: Query only
Options: Option R&S SMAB-K28
Manual operation: See " State - Gate " on page 238
:CALCulate[:POWer]:SWEep:TIME:GATE<ch>:STARt
<Start>
:CALCulate[:POWer]:SWEep:TIME:GATE<ch>:STOP <Stop>
Sets the start time of the selected gate. Insert value and unit.
Parameters:
<Stop> float
Increment: 1E-12
*RST: Start/Stop: 5/15 (Gate1), 25/35 (Gate2)
Example:
CALC:SWE:TIME:GATE2:STAR 20us
Sets a start time of 20 us for gate 2.
CALC:SWE:TIME:GATE2:STOP 30us
Sets a stop time of 30us for gate 2.
Options: Option R&S SMAB-K28
Manual operation: See " Start / Stop - Gate " on page 238
:CALCulate[:POWer]:SWEep:TIME:GATE<ch>:STATe <State>
Activates the gate settings for the selected trace. The measurement is started with
command SENS:POW:INIT. Both gates are active at one time.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
CALC:SWE:TIME:GATE:STAT ON
Enables time gated measurement.
Options: Option R&S SMAB-K28
Manual operation: See " State - Gate " on page 238
Power Sensor Measurement Subsystems