Remote Control Commands
R&S
®
SMA100B
468User Manual 1178.3834.02 ─ 05
:SENSe[:POWer]:SWEep:FREQuency:STOP <Stop>
Sets the stop frequency for the frequency mode.
Parameters:
<Stop> float
Range: 0 to 1E12
*RST: 22GHZ
Example:
SENS:SWE:FREQ:STOP 20.0GHZ
Sets the sweep stop to 20 GHz
Manual operation: See "Max - Power Analysis" on page 232
:SENSe[:POWer]:SWEep:FREQuency:TIMing[:MODE] <Mode>
Selects the mode in terms of speed and precision of the response of a measurement.
Parameters:
<Mode> FAST | NORMal | HPRecision
FAST
Selection FAST leads to a fast measurement with a short inte-
gration time for each measurement step.
NORMal
NORMal leads to a longer but more precise measurement due
to a higher integration time for each step.
*RST: FAST
Example:
SENS:SWE:FREQ:TIM:MODE FAST
The fast measurement mode is selected.
Manual operation: See " Timing - Power Analysis " on page 233
:SENSe[:POWer]:SWEep:FREQuency:YSCale:AUTO <Auto>
Activates autoscaling of the Y axis of the diagram.
Parameters:
<Auto> OFF | CEXPanding | FEXPanding | CFLoating | FFLoating
OFF
Auto scaling is deactivated. If switching from activated to deacti-
vated Auto scaling, the scaling is maintained.
CEXPanding | FEXPanding
Auto scale is activated. The scaling of the Y-axis is selected in
such a way, that the trace is always visible. To this end, the
range is expanded if the minimum or maximum values of the
trace move outside the current scale. The step width is 5 dB for
selection course and variable in the range of 0.2 db to 5 dB for
selection fine.
Power Sensor Measurement Subsystems