Remote Control Commands
R&S
®
SMA100B
630User Manual 1178.3834.02 ─ 05
0|OFF
Provided only for backward compatibility with other
Rohde & Schwarz signal generators.
The R&S SMA100B accepts these values and maps them auto-
matically as follows:
0|OFF = OFFTable
ONTable
Starts with the attenuation setting from the table and continues
with automatic level control.
*RST: AUTO
Example:
POW:ALC ON
Activates internal level control.
Manual operation: See "State" on page 198
[:SOURce<hw>]:POWer:ALC:DSENsitivity <Sensitivity>
Sets the sensitivity of the ALC detector.
Parameters:
<Sensitivity> AUTO | FIXed
AUTO
Selects the optimum sensitivity automatically.
FIXed
Fixes the internal level detector.
*RST: AUTO
Example:
POW:ALC:DSEN FIX
Manual operation: See "Detector Sensitivity" on page 198
[:SOURce<hw>]:POWer:ALC:SONCe
Activates level control for correction purposes temporarily.
Example:
POW:ALC OFF
Deactivates automatic level control at the RF output.
POW:ALC:SONC
Executes level control (once).
Usage: Event
Manual operation: See "Readjust" on page 71
[:SOURce<hw>]:POWer:ATTenuation:RFOFf:MODE <Mode>
Selects the state the attenuator is to assume if the RF signal is switched off.
Parameters:
<Mode> UNCHanged | FATTenuation
SOURce Subsystem