Remote Control Commands
R&S
®
SMA100B
639User Manual 1178.3834.02 ─ 05
[:SOURce]:ROSCillator:OUTPut:ALTernate:FREQuency:MODE..........................................642
[:SOURce]:ROSCillator[:INTernal]:ADJust:VALue.............................................................. 642
[:SOURce]:ROSCillator[:INTernal]:ADJust[:STATe].............................................................643
[:SOURce]:ROSCillator:PRESet
Resets the reference oscillator settings.
Example: See Example "Configuring the reference oscillator" on page 638.
Usage: Event
Manual operation: See "Set to Default" on page 273
[:SOURce]:ROSCillator:SOURce
<Source>
Selects between internal or external reference frequency.
Parameters:
<Source> INTernal | EXTernal
*RST: n.a. (factory preset: INTernal)
Example: See Example "Configuring the reference oscillator" on page 638.
Manual operation: See "Source" on page 273
[:SOURce]:ROSCillator:INTernal:TUNing[:STATe] <State>
Activates the EFC (external frequency control).
Parameters:
<State> 0 | 1 | OFF | ON
*RST: n.a. (factory preset: 1)
Example: See Example "Configuring the reference oscillator" on page 638.
Manual operation: See "External Tuning Active" on page 274
[:SOURce]:ROSCillator:INTernal:TUNing:SLOPe <State>
Sets the sensitivity of the external tuning volatge.
Parameters:
<State> LOW | HIGH
*RST: n.a. (factory preset: LOW)
Example: See Example "Configuring the reference oscillator" on page 638.
Manual operation: See "External Tuning Slope" on page 275
SOURce Subsystem