EasyManua.ls Logo

Rohde & Schwarz SMAB-B92 - Page 669

Rohde & Schwarz SMAB-B92
778 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMA100B
667User Manual 1178.3834.02 ─ 05
:SYSTem:COMMunicate:SERial:RESource?
Queries the visa resource string for the serial remote control interface. This string is
used for remote control of the instrument.
Return values:
<Resource> string
Example: See Example "Finding out the used VISA ressource strings"
on page 654.
Usage: Query only
:SYSTem:COMMunicate:SERial:SBITs
<SBits>
Defines the number of stop bits for the serial remote control interface.
Parameters:
<SBits> 1 | 2
*RST: n.a. (factory preset: 1)
Example: See Example "Finding out the used VISA ressource strings"
on page 654.
Manual operation: See "Stop Bits" on page 363
:SYSTem:COMMunicate:SOCKet:RESource?
Queries the visa resource string for remote control via LAN interface, using TCP/IP
socket protocol.
Return values:
<Resource> string
Example: See Example "Finding out the used VISA ressource strings"
on page 654.
Usage: Query only
:SYSTem:COMMunicate:USB:RESource?
Queries the visa resource string for remote control via the USB interface.
Return values:
<Resource> string
Example: See Example "Finding out the used VISA ressource strings"
on page 654.
Usage: Query only
:SYSTem:HELP:EXPort
Saves the online help as zip archive in the user directory.
SYSTem Subsystem

Table of Contents