EasyManua.ls Logo

SLE SLE5000 - Page 109

SLE SLE5000
292 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 109 of 292
Press the reset button to clear the alarm panel of the alarm notification.
Step 31 Reconnect the flow sensor and a new audible /
visual alarm Calibrate Flow Sensor should be
activated and replace the Connect Flow Sensor
alarm.
Recalibrate the flow sensor. Repeat Step 4 and then advance to Step 32.
Step 32 Press the Mode Select button to enter the Mode
select panel.
Step 33 For the SLE4000 ventilator: Select Ventilation Off from the Mode Select
panel. Press the confirm button. Advance to Step 42.
For the SLE5000 ventilator: Select HFO Only from the Mode Select panel, but
do not press the confirm button. Advance to Step 34.
SLE4000
SLE5000

Table of Contents

Related product manuals