Page 245 of 292
25. Bacterial Filters
It is recommended that bacterial filters are fitted in the fresh gas supply and on
the patient side of the exhalation block.
The filters reduce the possibility of infection to the patient and contamination of
the ventilator from secretions or fluids in the breathing circuits that could
accidentally enter the ventilators gas ports.
The ventilator can be used without bacterial filters in place, but the user must take
extra care in not allowing secretions or fluids to enter the ventilators gas ports.
25.1 Bacterial filter, SLE Part Nº:N2029 (Autoclavable)
This autoclavable bacterial filter is fitted into the humidifier supply line and has to
be fitted in accordance with the indicator arrow embossed on the surface of the
filter.
Warning: Do not immerse the filter in any liquid. If the filter is allowed to
become wet it will no longer function correctly and should be discarded.
Autoclave with pure dry saturated steam at:
134ºC (277ºF) (Allowable variation of temperature of +3ºC) at 220kPa (32psi) with
a minimum holding time of 3 minutes
or
121ºC (248ºF) (Allowable variation of temperature of +3ºC) at 96kPa (14.1psi)
with a minimum holding time of 15 minutes.
The filter can be autoclaved a maximum of 25 times within its anticipated service
life of 12 months.
For other makes of bacterial filter please refer to manufacturers instructions.
Bacterial Filter
Bacterial Filter
(SLE N
o
:N2587)
(SLE N
o
:N2029)