11.1.15 Functional Test of HFO Mode
Step 34 Set the following parameters in the HFO Only preview mode
HFO Rate.............10Hz
Mean ....................5mbar
FiO
2
......................21%
Delta P .................Not available in preview mode
After pressing the confirm button the user will be presented with the
HFO screen.
Set the Delta P to 20mbar
Step 35 Verify that:
the ventilator is oscillating,
the oscillation waveform appears in the waveform windows.