23.4 RS232 Connection Settings and Testing Data Output
Connect a standard serial cable from the ventilator to the test computer. Turn on
the ventilator and allow it to enter ventilation off mode. Open the Hyper Terminal
for Windows or the terminal emulator for other operating systems, (Settings:
COM1, 19200 bps, 8 Data bits, Parity None, 1 stop bit, Flow None or No
handshake). Turn on the ventilator, the hyper terminal should show that comma
delimited data is being sent.
100 Monitor/Display Comms Fail
101 EEProm Flow data corrupt
102 EEProm Oxygen data corrupt
103 EEProm Pressure Offset data corrupt
104 EEProm Pressure Gain data corrupt
105 EEProm Pressure Time constant Data corrupt